Move from a field vector to a coherence matrix
For a monochromatic transverse field write the complex amplitudes as \(\mathbf e=(E_x,E_y)^T\) with time convention \(e^{-i\omega t}\). A deterministic optical element gives
A detector averaging over unresolved time, frequency or spatial channels instead sees a coherence matrix \(C=\langle\mathbf e\mathbf e^\dagger\rangle\), which remains a \(2\times2\) Hermitian positive semidefinite matrix. A fixed Jones element acts as \(C_{\rm out}=JC_{\rm in}J^\dagger\). An incoherent mixture of distinct elements acts as \(\sum_a p_a J_a C_{\rm in}J_a^\dagger\), generally not representable by one common Jones matrix.
Use the convention \(S_0=C_{xx}+C_{yy}\), \(S_1=C_{xx}-C_{yy}\), \(S_2=2\Re C_{xy}\), \(S_3=2\Im C_{xy}\). This explicitly sets the sign of the circular component independent of competing right-handedness naming conventions. The determinant is \([S_0^2-S_1^2-S_2^2-S_3^2]/4\). Positive eigenvalues require \(P=\sqrt{S_1^2+S_2^2+S_3^2}/S_0\le1\) for nonzero signal. A pure Jones field gives rank one and \(P=1\); a mixture can give smaller \(P\). A fixed lossy polariser can convert an unpolarised input into a fully polarised transmitted field, so attenuation and loss of polarisation are separate observations.
Derive the Mueller matrix of a retarder and an unresolved mixture
Write \(C=\tfrac12\sum_j S_j\tau_j\), with \(\tau_0\) identity, \(\tau_1={\rm diag}(1,-1)\), \(\tau_2\) the real off-diagonal Pauli matrix, and \(\tau_3\) chosen with \(+i\) in its upper off-diagonal entry for the stipulated convention. Since \({\rm tr}(\tau_i\tau_j)=2\delta_{ij}\),
Inserting \(JCJ^\dagger\) gives the real Mueller matrix \(M_{ij}=\tfrac12{\rm tr}(\tau_i J\tau_jJ^\dagger)\). A common phase on \(J\) cancels. These intensity measurements cannot determine a spatially uniform optical carrier phase, even if every Mueller entry is known.
For \(J={\rm diag}(1,e^{i\delta})\), \(C_{xy}\) is multiplied by \(e^{-i\delta}\). Thus \(S_0,S_1\) remain fixed while \(S_2'=S_2\cos\delta+S_3\sin\delta\) and \(S_3'=-S_2\sin\delta+S_3\cos\delta\). A \(45^\circ\) linear input has \((S_1,S_2,S_3)/S_0=(0,1,0)\). A quarter-wave retardance sends it to \((0,0,-1)\) in the stated sign convention; a half-wave retardance sends it to \((0,-1,0)\). Measuring only horizontal and vertical powers would return equal readings for all three states. A diagonal analyser and a quarter-wave analyser are needed to distinguish them.
Now suppose half the unresolved beam spots undergo retardance \(+\delta\) and half \(-\delta\). Average their Mueller actions rather than averaging Jones fields. The \(S_2,S_3\) cross terms cancel and both components acquire a factor \(\cos\delta\). For \(\delta=\pi/3\), the diagonal linear input becomes \(P=0.50\), while a horizontal input remains fully polarised. This is anisotropic depolarisation. Resolving the two spots would reveal two fully polarised local states. A coherent recombination of amplitudes with a controlled phase is a different physical experiment and requires adding fields before averaging.
Reconstruct an element from independent incident states
Put four independent known incident Stokes vectors into columns of a matrix \(W\), and the corresponding measured output vectors into \(Y\). A linear optical response obeys \(Y=MW\), so \(M=YW^{-1}\) if \(W\) is nonsingular. Horizontal, vertical, diagonal-linear and a circular state with calibrated sign are one possible independent set. Four purely linear input states have \(S_3=0\), so their columns span at most a three-dimensional subspace. No amount of repetition of only those input states reveals the fourth response column. Additional independent preparations are more informative than merely increasing nominal data volume.
For normalised inputs \(H=(1,1,0,0)\), \(V=(1,-1,0,0)\), \(D=(1,0,1,0)\), \(C_+=(1,0,0,1)\), call measured outputs \(y_H,y_V,y_D,y_C\). The first column is \((y_H+y_V)/2\), the second \((y_H-y_V)/2\), the third \(y_D-M_{:0}\), and the fourth \(y_C-M_{:0}\). These differences assume known equal incident intensities or a correction to a common monitored intensity. A source that dims during one preparation can look like diattenuation. Measure an independent input monitor and repeat the preparations in a different order to distinguish drift from element properties.
Suppose the outputs are \((1,\pm1,0,0)\) for \(H,V\), \((1,0,0.50,0)\) for \(D\), and \((1,0,0,0.50)\) for \(C_+\). The reconstructed matrix is diagonal with entries \(1,1,0.50,0.50\), as in the unresolved equal-retardance example. The matrix establishes the action on Stokes states but does not uniquely identify that microscopic mechanism. A temporal fluctuator and an unresolved spatial mixture can give the same matrix. Change acquisition time, spectral bandwidth or spatial aperture to test which ensemble the detector is averaging.
Recover the retardance of a real plate with an imperfect analyser
Send a unit \(45^\circ\) linear input through the diagonal retarder above. Measuring the two independent output components gives
Thus \(\delta=\operatorname{atan2}(-S_3,S_2)\) modulo \(2\pi\). A fit using only the diagonal analyser, and hence only \(\cos\delta\), cannot distinguish opposite retardances and becomes insensitive near a maximum or minimum. The circular analyser supplies the missing quadrature. If the beam is partially depolarised equally in these two components, the angle from their ratio can remain well determined while their magnitude shrinks.
Suppose calibrated output readings give
with negligible \(S_1\). The inferred retardance is \(\operatorname{atan2}(0.64,0.48)=53.1^\circ\), while the degree of polarisation is \(0.80\). A single ideal Jones retarder acting on a pure input at one frequency could produce the angle but would preserve unit degree; something in the measured ensemble or instrument reduced it. A pure retarder model fitted only to \(S_2\) would instead infer \(\arccos0.48=61.3^\circ\), wrongly absorbing the contrast loss into its phase.
An analyser with leakage can also reduce the measured Stokes components. A differential pair of orthogonal polarisers with identical leakage fraction \(\eta\) scales its normalised difference by \((1-\eta)/(1+\eta)\). Calibrate each channel using known pure preparations and monitor intensity drifts before assigning all contrast loss to the specimen. Repeating with narrower spectral bandwidth or a smaller beam spot then tests the ensemble dependence of the inferred retardance and depolarisation.
Separate common optical phase from relative retardance
Inserting an isotropic plate can add the same phase to both orthogonal components of every coherent input. Its ideal Mueller matrix is unchanged because \(J\) and \(e^{i\alpha}J\) act identically on the coherence matrix. An interferometer with an external reference arm can nevertheless detect the added phase. This experiment supplies precisely the reference that a standalone polarimeter lacks. The two instruments can thus disagree about whether anything changed while measuring distinct physical observables accurately.
Tutorial on polarimetric inference
An analyser reports intensities \(I_H=0.60,\ I_V=0.40,\ I_D=0.75,\ I_A=0.25\) in common units. Which Stokes entries follow, and what remains unknown?
Work through the solution
The first pair gives
the diagonal pair gives \(S_2=0.50\). No circular analysis was performed, so \(S_3\) remains unknown within \(|S_3|\le\sqrt{1-0.20^2-0.50^2}=0.843\). Assuming a pure state would set its magnitude to that bound but would still not determine its sign. A partially polarised beam can have any smaller magnitude.
An ideal deterministic \(J={\rm diag}(\sqrt{0.8},\sqrt{0.2})\) transmits an unpolarised input with unit incident intensity. What are the output intensity and degree of polarisation?
Work through the solution
The input coherence matrix is half the identity. The transmitted diagonal entries are \(0.40\) and \(0.10\), so \(S_0=0.50,S_1=0.30\) and \(P=0.60\). The element selectively transmits one component of the mixture. It has not generated optical energy; half the incident intensity was lost.
A fitted Mueller matrix maps one measured positive input into \(P=1.08\). How should the inference be checked?
Work through the solution
That output violates positivity for this convention. Check dark subtraction, intensity normalisation, handedness of the circular analyser and conditioning of the preparation and analysis matrices. Collect more independent states with uncertainties and fit within a physical response model. Clipping each output degree to unity separately need not produce one common linear physical Mueller map for the instrument.
Reconstruct a Stokes state numerically from analyser intensities.
TU Delft's optical course covers polarisation states and their measurement.