Expand the recorded intensity and identify four terms
Let \(O(x)\) be the unknown coherent object field on the detector and \(R(x)=A_r e^{i2\pi f_cx}\) a known reference plane wave. They must have sufficient mutual coherence at the relative path delay, and their polarisation projections on the analyser must overlap. A square-law camera records \(I=|O|^2+|A_r|^2+O A_r^*e^{-i2\pi f_cx}+O^* A_re^{i2\pi f_cx}\). The first two terms are a real background and the last two are complex conjugates making real interference fringes. A local phase shift in \(O\) changes the fringe position without requiring the camera itself to follow the optical frequency.
Suppose the complex object spectrum has support approximately \(|f|\le B\). The spectrum of \(|O|^2\) is its autocorrelation and can extend to \(|f|=2B\). The cross terms occupy intervals centred at \(\pm f_c\) with half-width \(B\). In an ideal one-dimensional separation, \(f_c-B>2B\) keeps a sideband clear of the central intensity spectrum. A camera with pitch \(p\) has Nyquist frequency \(f_N=1/(2p)\), requiring \(f_c+B<f_N\). Both inequalities hold simultaneously only if \(B<f_N/4\) under these conservative support assumptions. Some fields have much narrower background or sparse spectra, but a designed full-bandwidth instrument cannot assume that without evidence.
Filter the negative-frequency sideband containing \(OA_r^*\), shift it to zero frequency, and inverse transform. Dividing by the known complex reference amplitude recovers the object field up to an overall calibration phase. Filtering the opposite sideband gives the conjugate field. A reference tilt calibration error leaves a linear phase ramp after reconstruction. Finite detector field of view multiplies the hologram by a window and spreads its Fourier peaks. Use a tapered window or leave guard bands at the cost of spatial resolution. The filter radius can suppress noise while also discarding high object frequencies; it is a physical resolution choice as well as a plotting parameter.
Choose a carrier that a real pixel lattice can record
Take a camera pitch \(p=6.5\,\mu{\rm m}\), giving \(f_N=76.9\) cycles per millimetre. If the field bandwidth is \(B=12\) cycles per millimetre, the conservative carrier interval is \(36<f_c<64.9\) cycles per millimetre. A choice \(f_c=48\) cycles per millimetre places the selected sideband between 36 and 60, clear of the central interval extending to 24. The fringe period is \(20.8\,\mu{\rm m}\), or 3.21 pixels. At vacuum wavelength \(532\,{\rm nm}\) and propagation close to normal in air the tilt angle is \(\arcsin(\lambda f_c)\approx1.46^\circ\). Raising the angle indefinitely would eventually alias the sideband back into the usable spectrum.
Each camera pixel integrates across its physical sensitive area. A full-width square pixel attenuates a sinusoidal fringe amplitude by \(\operatorname{sinc}(\pi f_c p)\), with \(\operatorname{sinc}x=\sin x/x\). At the example carrier this factor is about \(0.847\). Frequencies on either side of the band receive different attenuation, so a very broad sideband can have a frequency-dependent instrumental response even inside Nyquist. A flat-field intensity correction fixes pixel-to-pixel gain, but does not undo this finite-area transfer. The sensor's usable dynamic range and read noise also constrain the reference amplitude.
Convert a reconstructed phase into a path change
Let the reconstruction relative to a reference frame provide \(O_2O_1^*=|O_1O_2|e^{i\Delta\phi}\). The angle \(\arg(O_2O_1^*)\) gives phase change modulo \(2\pi\), so \(\Delta{\rm OPL}=\lambda\Delta\phi/(2\pi)\) modulo one vacuum wavelength. For transmission through a uniform nonabsorbing layer replacing ambient index \(n_a\), an added thickness \(d\) gives optical-path change \((n-n_a)d\). For a normally reflecting surface displacement \(h\) the outgoing and return paths both change, giving \(h=\lambda\Delta\phi/(4\pi)\) in air. Using the transmission conversion on reflection doubles inferred height.
A reflection measurement at \(532\,{\rm nm}\) gives a local phase change \(0.80\,{\rm rad}\), implying a height change \(33.9\,{\rm nm}\) modulo \(266\,{\rm nm}\) under the ideal normal-incidence geometry. In transmission through a layer of contrast \(n-n_a=0.10\), the same phase corresponds to \(0.677\,\mu{\rm m}\) of thickness modulo \(5.32\,\mu{\rm m}\). A single phase image cannot simultaneously determine thickness and refractive index without independent information. Phase unwrapping uses spatial or temporal continuity to choose integer cycles; at dark amplitude nodes or steep unresolved steps those choices can be ambiguous.
Background · How can a second wavelength extend the unambiguous path range?
Reconstruct the phase of the same optical path change \(d\) at two vacuum wavelengths. Assume the path change is independent of wavelength and both reconstructions share a calibrated zero. Each phase is known modulo \(2\pi\). Their difference increases more slowly with path than either phase alone.
Wavelengths of 630 and 640 nm give a synthetic path period of 40.32 µm. For normal reflection the physical surface displacement is half the round-trip path change. A 0.05 rad uncertainty in the difference phase corresponds to about 0.32 µm of path uncertainty at this synthetic period. The long period helps identify a fringe order, while a single-wavelength phase can refine the displacement once that order is known.
Measure noise and reference aberration separately
The fringe visibility for uniform signal and reference intensities is \(2\sqrt{I_oI_r}/(I_o+I_r)\) when coherence and polarisation overlap are perfect. Equal intensities maximise visibility. Increasing reference power can raise heterodyne signal above camera read noise, but once reference shot noise dominates, continually increasing it does not improve sensitivity proportionally and can saturate pixels. A fixed total photon budget and a fixed object photon budget are different optimisation problems. Background subtraction should use a reference-only frame, an object-only frame when feasible, and a blank complex phase calibration acquired with the same optical geometry.
A wavefront aberration in the reference multiplies the recovered object by a spatially varying phase. Dividing the complex field by a blank-field reconstruction can remove a stable aberration, but division near very low blank amplitude amplifies noise. A phase map through masked pixels should disclose where phase was interpolated and where the signal actually supported it. Reconstruct at several sideband filter radii and carrier angles to check whether a fine feature moves or changes shape with the processing. A real specimen feature should remain in specimen coordinates under those controlled changes within the valid resolution.
Decide whether a thin specimen can be measured without unwrapping
If surface steps in reflection are known a priori to lie within a single interval of width \(\lambda/2\), the wrapped phase can be mapped uniquely into that chosen range up to a common reference height. For \(532\,{\rm nm}\) a known \(60\,{\rm nm}\) relief is safely smaller than the \(266\,{\rm nm}\) height period. An isolated \(300\,{\rm nm}\) step can instead appear as \(34\,{\rm nm}\) if there is no continuity information across its edge. A smoothness algorithm can select precisely the wrong branch at a real discontinuity.
Two wavelengths offer a different reference scale. For nearby wavelengths \(\lambda_1,\lambda_2\), subtract their interferometric phases for the same optical path to obtain a synthetic optical-path period \(\lambda_1\lambda_2/|\lambda_1-\lambda_2|\). Reflection height has half that period. With \(532\) and \(540\,{\rm nm}\), the height period is about \(18.0\,\mu{\rm m}\), much larger than for either colour alone. The corresponding height sensitivity to a given phase noise is worse, so the synthetic phase often chooses the coarse branch while one individual wavelength supplies fine precision. This assumes both measurements interrogate the same position and calibrated path without a significant wavelength-dependent sample response.
Tutorial on carrier and phase measurements
The same \(6.5\,\mu{\rm m}\) camera is asked to recover \(B=22\) cycles per millimetre with full separation from an object intensity spectrum extending to \(2B\). Is there any carrier satisfying both conditions?
Work through the solution
The carrier would have to exceed \(66\) and be below \(76.9-22=54.9\) cycles per millimetre. No such carrier exists. Change magnification to reduce object bandwidth on the detector, use smaller pixels, or exploit multiple phase-shifted frames under an appropriate measurement model. Merely tuning the tilt cannot create detector bandwidth.
A reconstruction of a flat mirror shows an apparent phase ramp of \(2\pi\) across \(5\,{\rm mm}\) after demodulation. What carrier correction would remove a pure calibration ramp?
Work through the solution
A phase change of one cycle over \(5\,{\rm mm}\) corresponds to \(0.20\) cycles per millimetre. Adjust the assumed carrier by that magnitude with sign selected from the observed phase slope, then repeat on the reference mirror. Correcting a sample ramp without a blank could accidentally remove a real tilt of the sample.
A wrapped phase map appears very smooth even though the object amplitude almost vanishes on one side of an image. Does smoothness certify accuracy?
Work through the solution
Near zero complex amplitude the phase responds very strongly to noise. Smoothing or interpolation can produce a visually regular map without adding phase information. Examine the sideband signal-to-noise ratio, repeat at independent exposures, and state an amplitude-based confidence mask. An optical-path conversion is quantitative only in the region that supplied a reliably measured complex field.
Cuche, Marquet and Depeursinge demonstrate Fourier separation of off-axis holographic orders.
Vary a Fourier aperture and observe the resolution cost of filtering.