Fundamentals / Physics / Waves and optics / Optical fields, instruments and observations

What an objective must collect to resolve a specimen

A microscope can transmit plenty of light and still erase fine detail. Treat the specimen as a weak grating, follow its spatial harmonics through the objective, and determine why changing illumination coherence changes the transfer of contrast.

Diffract a sinusoidal object into orders

Use a thin specimen with amplitude transmission \(t(x)=1+\epsilon\cos(2\pi fx)\), \(|\epsilon|\ll1\), and uniform normally incident coherent light. Write the cosine as half the sum of two exponentials. Immediately after the sample the field has spatial frequencies zero and \(\pm f\). A lens maps transverse spatial frequency into angle, and its finite numerical aperture \({\rm NA}=n\sin\theta_{\max}\) accepts \(|f|\le{\rm NA}/\lambda_0\) for coherent normal illumination, where \(\lambda_0\) is the vacuum wavelength. Beyond that limit both first orders are blocked, leaving only a constant image field and no sinusoidal contrast.

If the zero order and both sidebands pass through a clear ideal pupil, inverse Fourier transformation reproduces \(1+\epsilon\cos(2\pi fx)\). Its intensity is approximately \(1+2\epsilon\cos(2\pi fx)\). If only one sideband and the zero order pass, their interference still produces a feature at \(f\), but the first-order intensity contrast is halved for this object and the image field acquires a spatially varying phase. If the zero order is suppressed as well, the intensity contrast follows a different mixing law. These possibilities show why the statement that one needs two interfering orders must specify which orders and which image quantity is measured.

Oblique coherent illumination shifts the object's spatial spectrum by the incident transverse frequency \(f_i\). A sideband initially outside the pupil can enter while one from the opposite direction exits. With illumination aperture \({\rm NA}_{\rm illum}\) and objective aperture \({\rm NA}_{\rm obj}\), the extreme accessible object frequency from angular coverage can reach \(({\rm NA}_{\rm illum}+{\rm NA}_{\rm obj})/\lambda_0\). That support limit alone does not guarantee uniform contrast or isotropic information. The illumination distribution and reconstruction determine how well any frequency inside the support can be measured.

Derive the incoherent transfer from a pupil autocorrelation

For spatially incoherent emitters, intensities from different object points add. If \(h\) is the coherent amplitude point-spread function, the incoherent intensity point-spread function is \(|h|^2\), and the intensity image is its convolution with object brightness. Fourier transforming the product \(|h|^2\) produces an autocorrelation of the pupil \(P\). A clear circular pupil gives a transfer value proportional to the overlap of two equal disks displaced by a spatial-frequency-dependent distance. At zero shift the disks overlap fully; at a diameter of shift their overlap vanishes. The incoherent cutoff is therefore \(f_c=2{\rm NA}/\lambda_0\), twice the coherent on-axis field cutoff.

For \(u=|f|/f_c\le1\), the normalised overlap area is

\[{\rm MTF}(u)=\frac{2}{\pi}\left[\cos^{-1}u-u\sqrt{1-u^2}\right].\]

To derive it divide the overlap lens into two symmetric disk segments. Each segment is a circular sector of angle \(2\arccos u\) with its isosceles triangle subtracted. Summing and dividing by the full area yields the expression above. It declines continuously and gives about \(0.391\) at half the incoherent cutoff. The coherent transfer of a clear pupil has unit field transmission inside its support; it operates on complex amplitude. These two curves describe different imaging experiments. Replacing a coherent field by its intensity before applying a coherent pupil would use the wrong linear system.

For \(\lambda_0=0.550\,\mu{\rm m}\) and \({\rm NA}=0.80\), the coherent cutoff is \(1.45\) cycles per micrometre and the incoherent cutoff \(2.91\) cycles per micrometre. An incoherent sinusoidal fluorescent brightness pattern at \(f=2.0\) cycles per micrometre has \(u=0.6875\) and ideal transmitted contrast about \(0.20\) of its original. A perfectly transmitting objective cannot recover the missing higher frequencies merely by magnifying the image farther. Magnification can, however, change camera sampling and therefore recover detail that the objective passed but undersampled camera pixels discarded.

Separate diffraction, sampling and noise in a resolution estimate

A detector with physical pixel pitch \(p\) at magnification \(M\) samples the specimen at \(p/M\). The Nyquist frequency is \(M/(2p)\). For \(p=6.5\,\mu{\rm m}\), \(M=40\) this gives \(3.08\) cycles per micrometre, just above the illustrative incoherent cutoff. Finite pixel area also averages over a square patch and multiplies the axial-frequency contrast by a sinc factor \(\sin(\pi f p/M)/(\pi f p/M)\). At \(2.0\) cycles per micrometre that factor is about \(0.835\). The optical and pixel factors multiply, leaving less measured contrast than an optical-only transfer curve predicts. A sinusoidal target in a diagonal orientation sees the product of the factors in the two camera axes.

A cutoff is a support statement. Discrimination in real noise requires a contrast or likelihood criterion, a measured background and a known photon count. Deconvolving a frequency where the transfer is \(0.1\) multiplies noise in that component by roughly ten when dividing by the transfer. At a true zero no such division can restore the missing datum without additional assumptions or measurements. A point-pair Rayleigh separation, the highest sinusoidal frequency transmitted, and precision in locating the centre of an isolated bright emitter answer three different quantitative questions.

Compare the pupil models directly

The live graph plots the normalised coherent field pupil and incoherent intensity modulation transfer on the same specimen-frequency axis, with a separately identified pixel-averaged incoherent curve. Change numerical aperture while holding camera sampling fixed, then change wavelength. Predict whether optical cutoff or camera Nyquist is limiting before reading the returned values. The rectangular coherent curve should be interpreted as amplitude transmission for normally incident coherent light, whereas the smoother curves act on brightness contrast of incoherent emitters.

Measure transfer with a sinusoidal calibration target

For an incoherent test pattern of mean brightness \(I_0\) and intrinsic modulation \(m_o\), measure the image modulation as \(m_i=(I_{\max}-I_{\min})/(I_{\max}+I_{\min})\) after correcting the additive detector background. In a linear shift-invariant imaging regime the ratio \(m_i/m_o\) estimates the magnitude of the total transfer at the target frequency. A constant stray background reduces the measured modulation even if the pupil is unchanged. It increases the denominator without increasing the numerator, so subtract its independently measured intensity before comparing with a diffraction formula.

For a target of modulation \(0.60\) producing corrected maximum and minimum signals \(550\) and \(450\) counts, the image modulation is \(0.10\), so the inferred transfer is \(0.167\). The earlier optical and pixel factors at two cycles per micrometre predict approximately this value. A target with sharp bars contains many harmonics rather than one pure frequency. Its peak-to-trough contrast cannot generally be inserted into the same single-frequency formula without accounting for those harmonics. A sinusoidal target or a measured Fourier coefficient gives a cleaner comparison with the pupil calculation.

Compare diffraction conventions across the pupil model

When checking a spatial transfer against a diffraction calculation, keep the vacuum wavelength, refractive index and numerical aperture together: index already enters through numerical aperture. Inserting the wavelength in the medium a second time in a formula written with vacuum wavelength would count that index twice. The distinction matters most for immersion objectives and for comparisons of cutoffs reported from different beam geometries.

The open TU Delft diffraction chapter develops the wave-optical treatment of apertures and optical systems.

Tutorial on resolution claims

A normal-incidence coherent image of a grating with period \(0.50\,\mu{\rm m}\) is uniform under the example objective. Can changing only illumination angle help?

Work through the solution

The spatial frequency is \(2.0\) cycles per micrometre, beyond the on-axis coherent limit \(1.45\). An illumination shift of order \(0.55\) cycles per micrometre can bring one first order inside while retaining the reference order, restoring some contrast. The shift corresponds to illumination NA about \(0.30\) at the stated wavelength. Other orientations can respond differently.

At \(20\times\) magnification with the same camera, which limit applies to an incoherent \(2.0\)-cycle-per-micrometre pattern?

Work through the solution

The specimen-plane sampling pitch is \(0.325\,\mu{\rm m}\), giving Nyquist \(1.54\) cycles per micrometre. Optics can transmit the pattern, but the camera aliases it. Increasing magnification can help until sampling ceases to be limiting; it cannot raise the optical transfer at a fixed specimen frequency.

In an ideal coherent microscope a weak pure-phase grating has \(t(x)\approx1+i\epsilon\cos(2\pi fx)\). Does passing all three orders guarantee first-order intensity contrast in focus?

Work through the solution

No first-order intensity term survives because the perturbation is in quadrature with the reference. The field modulation is present but the intensity changes only at higher order in this approximation. Introduce a controlled reference phase shift, defocus, or another phase-sensitive measurement to convert the phase modulation into brightness contrast. Distinguish absence of an order from absence of interference in the detected quadrature.

TU Delft's Advanced Optical Imaging course derives coherent image formation from the pupil.