Count optical states in a large equilibrium cavity
Take a cubic airless cavity of volume \(V\) in thermal equilibrium at temperature \(T\). For its bulk mode density we can use periodic boundary conditions. Allowed wavevectors are separated by \(2\pi/L\) in each coordinate. Count a spherical shell in wavevector space, divide by the volume per state and include two independent transverse polarisations. Substituting \(k=2\pi\nu/c\) gives
Using standing waves in an ideally reflecting large box gives the same leading volume density when directions and polarisations are counted consistently.
One oscillator mode of frequency \(\nu\) can hold \(m=0,1,2,\ldots\) photons. Its thermal occupation has probabilities proportional to \(e^{-mh\nu/k_BT}\). Summing the geometric series and taking its logarithmic derivative gives
Multiplying by the density of modes and energy per photon gives spectral energy density \(u_\nu\). The zero-point contribution of a harmonic oscillator is not part of the thermal power exchanged between these black surfaces. At low frequency the result approaches an energy \(k_BT\) per mode; extending that classical limit to arbitrarily high frequency would make the integrated power diverge.
An isotropic distribution of energy travelling at speed \(c\) produces a normal flux through a small aperture weighted by the direction cosine. Per unit solid angle its radiance is \(B_\nu=cu_\nu/(4\pi)\). Transform to equal wavelength intervals using \(B_\lambda d\lambda=B_\nu|d\nu|\). The two useful spectral forms are
The wavelength expression has units of watts per square metre per steradian per metre of wavelength. To express it per micrometre multiply by one millionth. At 400 K and 10 micrometres the result is 33.56 W per square metre per steradian per micrometre. At 295 K it is 9.14 in the same units. Equal wavelength and equal frequency bins carry different widths, so their distributions reach their maxima at different converted wavelengths. A spectrum's peak should always name its horizontal variable and spectral-density convention.
Distinguish emitter power from brightness in a chosen direction
For a black surface with angle-independent radiance into its outward hemisphere, integrate \(\cos\theta\,d\Omega\). The result is \(\pi\), not \(2\pi\). The emitted spectral power per unit area is \(\pi B_\lambda\). Integrate over frequency using \(x=h\nu/(k_BT)\). The remaining integral \(\int_0^\infty x^3/(e^x-1)\,dx=\pi^4/15\) follows by expanding \(1/(e^x-1)\) as a sum of exponentials and integrating term by term. This yields Stefan–Boltzmann exitance \(M=\sigma T^4\) and identifies the constant \(\sigma=2\pi^5k_B^4/(15h^3c^2)\).
A black patch of 1.0 square millimetre at 400 K emits about 1.45 mW into its entire outward hemisphere over all wavelengths. The same patch at 295 K emits about 0.429 mW. Replacing one by the other changes all-wavelength hemispherical power by about 1.02 mW. This is not the power a small infrared receiver sees through a narrow filter. The integrated emitter number is valuable for energy conservation, but by itself contains no receiver area, direction, passband or optical transmission.
The surface spectrum of a body in thermal equilibrium has directional spectral emissivity equal to its absorptivity in the corresponding reciprocal optical modes under the usual reciprocal-material conditions. An opaque diffuse grey surface with emissivity \(\epsilon\) in an isotropically illuminated background can be modelled by \(\epsilon B_\lambda(T)+(1-\epsilon)B_\lambda(T_{\rm bg})\). It emits less than a black surface but also reflects background. A shiny object viewed in a warm room may consequently display a radiance temperature nearer the reflected scene than its own physical temperature. Specular surfaces require explicit directional knowledge of that scene.
Derive the geometric acceptance without counting the angle twice
An area element \(dA\) normal to the collection axis emits into \(d\Omega\) at angle \(\theta\) a spectral power \(B_\lambda\cos\theta\,dA\,d\Omega\). For a uniformly filled cone of half-angle \(\alpha\) and a uniform patch of area \(A\), integrating gives \(A\pi\sin^2\alpha\). This is its geometrical throughput in air when there is no vignetting. Optical étendue more generally uses \(n^2dA\cos\theta\,d\Omega\). A lossless passive optical relay in geometrical optics preserves this phase-space measure; correspondingly \(L/n^2\) is conserved along a ray in transparent media under the usual stationary conditions.
A direct lens example makes the restriction tangible. An image relay enlarging a uniformly bright patch by a factor of two in each transverse direction forms an image four times the area. In a lossless paraxial design filling the original pupil, each image point then occupies one quarter the solid angle. A detector covering the whole enlarged image and accepting that cone can receive the original power. Placing a fixed small sensor at the image plane intercepts less area instead. Higher magnification has not created additional radiance, even if it makes a small detail easier to resolve.
For the same 1.0 square millimetre patch seen within ten degrees, the projected acceptance is \(9.47\times10^{-8}\) square metre steradian. A wavelength-independent combined transmission and detector absorption of 0.65 leaves effective acceptance \(6.16\times10^{-8}\) square metre steradian. Integrating the Planck curve from 8 to 12 micrometres gives source and background band radiances 133.74 and 35.43 W per square metre per steradian. Their difference yields only 6.05 microwatts in the receiver. Keeping a single midpoint radiance times four micrometres happens to be fairly close here, but there is no fixed universal midpoint correction across temperatures or filters.
Set source and ambient temperatures equal. The emitted and ambient curves remain finite while the differenced instrument reading vanishes. Increase the cone half-angle from ten to twenty degrees and predict the scale factor \(\sin^2 20^\circ/\sin^2 10^\circ=3.88\). The spectral shape should stay fixed if transmission is separable from angle. If real data shift spectral shape with aperture, inspect filter angle dependence, source nonuniformity and stray light. The integral in the readout is computed on the specified 8–12 micrometre band even though the graph shows wavelengths beyond it.
Convert received power to calibrated voltage or counts
A thermal detector may respond to absorbed power through a temperature rise and an electrical responsivity. A photon-counting detector measures absorbed photons, so its count rate contains a factor \(\lambda/(hc)\) multiplying spectral power before integration. Two receivers with identical geometric acceptance and nominal quantum efficiency can respond differently to a changing source temperature if one output is proportional to energy and the other to photon number. Calibrate the actual electrical transfer and make the time interval explicit for accumulated counts.
For an ideal linear power detector let \(V=V_0+gP\), with offset \(V_0\) and gain \(g\). Two calibrated sources spanning the use range determine both. A black reference port at the same ambient temperature can also reject a common background in a chopped measurement. A filter, window or warm lens that absorbs thermal radiation also emits it. If these optics remain unchanged and stable, part of their own emission cancels in a reference difference. A reference traversing a different path generally leaves a residual input that is not captured by an electronic dark measurement.
Suppose effective source-minus-ambient powers of 2.00 and 8.00 microwatts yield 0.120 and 0.420 V. The fitted slope is 0.0500 V per microwatt with an offset of 0.020 V. A reading of 0.320 V implies 6.00 microwatts after offset correction. Substituting the same reading directly into a single-point gain with zero assumed offset would bias the inferred temperature. Rechecking a third reference near six microwatts tests linearity, while repeated zero-input readings at realistic instrument temperature test offset drift.
Invert the radiance and propagate uncertainties
A narrowband blackbody radiance measurement can be inverted analytically by solving Planck's expression for \(T\). In a broad passband use numerical quadrature across the calibrated spectral response and solve a scalar monotone equation when emissivity and background are known. The derivative at a single wavelength is particularly informative for estimating precision. With \(x=hc/(\lambda k_BT)\),
At 10 micrometres and 400 K the logarithmic sensitivity is 3.70. A one-percent relative radiance uncertainty therefore gives about 1.08 K temperature uncertainty if all other quantities are exact. An unknown five-percent emissivity error can dominate that estimate. A two-colour temperature ratio may reduce uncertainty from an emissivity that is constant across the two narrow bands, but cannot cancel an arbitrary spectral emissivity or a large reflected ambient component.
As an example consider a grey object at 400 K and ambient 295 K with emissivity 0.70. At ten micrometres its outgoing radiance is \(0.70(33.56)+0.30(9.14)=26.24\) in the earlier spectral units. Inverting as a perfect blackbody gives a brightness temperature of about 375 K. The difference is a physical radiance–temperature degeneracy. Repeat at known ambient temperatures while keeping the object thermally stabilised to constrain the reflected fraction. If heating the background changes the actual object temperature, a second independent temperature sensor or a transient thermal model is necessary.
When detector heat capacity sets the measured signal
Consider a thermal receiver of heat capacity \(C_d\) coupled to its surroundings with thermal conductance \(G_d\). For small excursions its temperature obeys \(C_d\dot\theta+G_d\theta=P_{\rm abs}(t)\), where \(\theta\) is the excess above its operating point. A voltage thermometer with responsivity \(r_T\) converts this into output. Its frequency-dependent power responsivity is \(r_T/(G_d-i\omega C_d)\). The geometric and spectral calculations determine absorbed optical power, while this thermal equation determines how much of its modulation appears at the electrical output.
Take \(C_d=2.0\) microjoules per kelvin and \(G_d=0.10\) milliwatt per kelvin. The thermal time constant is 20 ms and ordinary corner frequency is 7.96 Hz. A sinusoidal absorbed-power amplitude of 6.0 microwatts would give a static temperature rise of 0.060 K. At 20 Hz the amplitude is reduced by \(1/\sqrt{1+(2\pi f\tau)^2}=0.370\), leaving about 22.2 mK with a phase lag magnitude near 68 degrees. A receiver calibrated with a steady optical power is not automatically calibrated for this chopped experiment.
A square-wave chopper also changes the relation between time-domain peak differences and a lock-in's fundamental reading. For equal hot and reference dwell times, a square wave switching between powers \(P_1,P_0\) has fundamental sinusoidal peak amplitude \(2(P_1-P_0)/\pi\), with different factors for rms and peak reports. The thermal transfer attenuates this component at the chopper frequency and the odd harmonics at their respective frequencies. Substituting the full hot-minus-reference power directly into the first-harmonic response would overestimate the expected signal.
Measure a frequency sweep on a stable reference and test whether amplitude and phase share a single time constant. A geometrical error rescales the amplitude but ordinarily leaves this thermal phase unchanged. A second thermal mass or an electronic filter can introduce another pole. Lowering chop frequency increases thermal signal until the static limit but can expose low-frequency environmental drift. Choose the frequency by comparing the calibrated response with the measured noise spectrum.
Finally, a reference and an unknown source must illuminate the same active area if responsivity varies across the receiver. Imaging one source as a small spot and the other as a filled patch can create a calibration bias even when total spectral powers agree. Translating a reference spot across the detector or imaging the same aperture successively at both temperatures tests that spatial uniformity assumption.
If the sensor is used on a source much smaller than the calibrated reference aperture, record an image or a field-stop scan to determine its filling fraction. Otherwise even exact knowledge of filter transmission and thermodynamic temperature leaves an unknown geometrical scale factor. Aperture filling is part of the measurement equation, not an electronic gain setting.
Tutorial · Infer emissivity using two controlled backgrounds
An opaque diffuse sample at fixed temperature is observed at 10 micrometres against two isotropic backgrounds with calibrated spectral radiances 8.00 and 16.00 W per square metre per steradian per micrometre. The resulting outgoing readings are 25.00 and 27.00 in the same units. Infer emissivity and the sample's blackbody-equivalent radiance. Estimate the significance of independent reading errors of 0.10 units.
Separate the reflected and emitted contributions
The increased outgoing radiance is 2.00 for an 8.00 increase of background. Thus reflectivity is 0.250 and emissivity 0.750. The first measurement gives intrinsic blackbody radiance \([25.00-0.25(8.00)]/0.75=30.67\). The second returns the same value. With 0.10 independent uncertainty on each outgoing reading and exact backgrounds, the slope uncertainty is \(\sqrt2(0.10)/8=0.0177\), as is the absolute emissivity uncertainty. Larger background separation can improve this slope estimate if sample temperature stays fixed.
If the sample transmits radiation, the residual fraction does not all correspond to reflection. A controlled change behind the sample helps detect that additional path. A directional specular reflector exposed to a warm enclosure also does not necessarily see the same angular average used in the diffuse model. Without these checks, a formally precise fitted emissivity can describe the experimental surroundings instead of the material.
Tutorial · Design a numerical spectral-integration check
An engineer tabulates spectral radiance per micrometre on a wavelength mesh. Another rewrites the detector model on an equally spaced frequency mesh, obtaining a signal that differs by several percent after independent mesh refinement. Explain what must transform, what stays dimensionless and how to compare a photon-counting channel with a power channel.
Transform the complete measurement integral
The differential power \(B_\lambda\,d\lambda\) must equal \(B_\nu\,|d\nu|\), with \(|d\lambda/d\nu|=c/\nu^2\). A dimensionless transmission curve becomes the same physical transmission at \(\lambda=c/\nu\); it acquires no additional Jacobian. The integration bounds reverse when increasing wavelength is replaced by increasing frequency. A response specified per unit wavelength, or a calibrated gain integrating a particular finite bin, must instead be interpreted from how it was defined.
Evaluate a unit-transmission test using a very narrow band, then the full 8–12 micrometre band whose radiance values are given above. Refine each grid independently, treating sharp filter edges consistently and converting per-micrometre densities before using metres. For a photon counter divide each spectral power contribution by photon energy before summation. Replacing that energy by a single central-wavelength value is an approximation that should be checked against the full band and source temperature. Close agreement between two different meshes using the same incorrect density is not a valid units check.
NIST radiance-temperature calibrations · spectral radiance, sources and uncertainty