Fundamentals / Physics / Electrodynamics / Electromechanics, optics and measurement

Optical trapping and Brownian calibration of a dielectric bead

An image of a bead remaining near a focus does not establish the strength of its trap. Derive the force pulling a small dielectric sphere into a bright region, compare it with the forward force from scattering and learn what a motion record can tell you about stiffness, drag and temperature. The model here uses an inert dielectric sphere in a homogeneous Newtonian liquid.

From two dielectric boundaries to the energy of an induced dipole

A sphere of radius \(a\) and relative permittivity \(\epsilon_b\) sits in a liquid of relative permittivity \(\epsilon_m\). Assuming the sphere is small compared with a wavelength in the liquid, solve its local electric problem in an almost uniform applied field \(E\). Rotational symmetry about the field admits an interior potential proportional to \(r\cos\theta\). The exterior consists of the imposed potential \(-Er\cos\theta\) and an induced dipolar term proportional to \(\cos\theta/r^2\). Equality of potential and of normal displacement at the surface gives the local-field and dipole factors

\[E_{\rm in}=\frac{3\epsilon_m}{\epsilon_b+2\epsilon_m}E,\]
\[p=\alpha E,\]
\[\alpha=4\pi\epsilon_0\epsilon_m a^3\frac{\epsilon_b-\epsilon_m}{\epsilon_b+2\epsilon_m}.\]

The field doing the polarising work is the external field in the host at the particle position before inserting it. To turn that work into a potential, build up the dipole from zero rather than assuming a permanent dipole moment. Quasistatic assembly gives

\[-\int_0^E\alpha E'\,dE'=-\alpha E^2/2.\]

For a harmonic peak field average the square over an optical period. In a nearly lossless liquid of refractive index \(n_m\), the peak field is related to travelling-wave intensity by \(I=n_m\epsilon_0c|E|^2/2\). The resulting conservative force is

\[U=-\frac{\operatorname{Re}\alpha}{4}|E|^2=-\frac{\operatorname{Re}\alpha}{2n_m\epsilon_0 c}I,\]
\[\mathbf F_{\rm grad}=-\nabla U .\]

The two factors of two have different origins: induced-dipole assembly and time averaging of a sinusoid. Using a permanent-dipole potential with a peak field would miss both. A bead of higher refractive index than the liquid has positive real polarizability and is attracted to increasing intensity. A lower-index inclusion reverses this local tendency. For large particles, strong absorption or rapidly structured fields, a dipole with real static polarizability no longer specifies the force; a field or multipole calculation of the complete momentum balance is required.

Investigate the small-sphere dipole response and its dependence on particle size

Compute a transverse stiffness and test how local it is

Assume the beam near its focal plane has intensity \(I(x)=I_0e^{-2x^2/w^2}\), with power \(P\), waist \(w\) defined at \(1/e^2\) intensity and \(I_0=2P/(\pi w^2)\). Put \(U_0=\operatorname{Re}\alpha I_0/(2n_m\epsilon_0c)>0\). The potential measured above its minimum is \(U_0[1-e^{-2x^2/w^2}]\). Expanding to second order and equating its leading term to \(\kappa_xx^2/2\) gives

\[\kappa_x=4U_0/w^2.\]

Stiffness therefore scales linearly with power, as the third power of sphere radius and as the inverse fourth power of waist at fixed indices. Each scaling applies within the stated approximations.

For a 40 nm radius bead with \(n_b=1.59\), take \(n_m=1.33\), \(w=0.65\) micrometres, vacuum wavelength 1064 nm, \(P=90\) mW and \(T=298\) K. The index contrast factor is 0.1252 and the waist intensity is \(1.36\times10^{11}\) watts per square metre. The predicted depth \(3.03\times10^{-20}\) J is 7.36 times the thermal energy. The lateral stiffness is \(2.87\times10^{-7}\) N/m. A displacement of 50 nm then requires approximately 14.3 fN of transverse balancing force. At a displacement of half a waist, the exact restoring force is the harmonic estimate multiplied by \(e^{-1/2}=0.607\); a Hookean calibration cannot be extrapolated that far.

Use the two curves to identify the interval in which the local quadratic energy is within ten percent of the Gaussian prediction. First lower power at fixed waist. Both curves shrink together in energy units while the harmonic thermal displacement grows. Next increase bead radius at fixed power and watch the well deepen, but also evaluate the size parameter \(2\pi n_ma/\lambda_0\). It is 0.314 at the default and grows linearly with radius. Stronger trapping predicted by this simplified model does not establish the accuracy of the dipole approximation at arbitrary sizes.

Compare axial restoring force with radiation pressure

An induced dipole reradiates a portion of the incident field. In a uniform host use the Rayleigh scattering cross section \(\sigma_s=(8\pi/3)k_m^4a^6|(\epsilon_b-\epsilon_m)/(\epsilon_b+2\epsilon_m)|^2\). In a common travelling-wave momentum-flux model for a bead in a stationary nondispersive liquid the axial scattering force is approximately \(F_s=n_m I\sigma_s/c\); absorption would add a further contribution. This estimate uses the total scattering cross section because the leading dipole pattern has vanishing mean forward cosine. For a larger forward-scattering sphere a transport cross section weights the deflected momentum instead.

In the default case the cross section is roughly \(2.05\times10^{-18}\) square metres. The forward force at the waist is about 1.23 fN. A Gaussian axial profile has \(I(0,z)=I_0/[1+(z/z_R)^2]\) with \(z_R=\pi n_mw^2/\lambda_0=1.66\) micrometres. Its local axial stiffness \(2U_0/z_R^2=2.20\times10^{-8}\) N/m is much smaller than its transverse stiffness. Balancing the constant forward force with that local stiffness predicts an equilibrium about 56 nm downstream of the waist. The small displacement relative to the Rayleigh length supports the linearisation.

Both the axial stiffness and forward scattering force scale with power in this simple model. A power sweep can thus increase confinement while barely shifting the equilibrium offset, a useful discriminating prediction. At fixed focus the offset grows rapidly with bead size because the scattering force scales with the sixth power and the gradient force with the third power. A stable lateral position alone does not establish stable three-dimensional capture. Check the axial restoring slope at a force zero and verify that thermal escape and fluid flow are negligible over the acquisition interval.

Extract stiffness from fluctuations or from a relaxation time

Near a stable point in one Cartesian direction, let \(x\) denote displacement from the mean and \(\zeta\) the viscous drag. At sufficiently low frequencies the inertial term is negligible and the overdamped Langevin equation is

\[\zeta\dot x=-\kappa x+\xi.\]

At equilibrium the fluctuating force has zero mean and white covariance \(2k_BT\zeta\delta(t-t')\) over the frequencies considered. The deterministic relaxation time is

\[\tau=\zeta/\kappa.\]

Multiplying the equation at positive delay by \(x(0)\) and averaging removes the uncorrelated later noise and gives the decaying covariance. Equipartition sets its initial value.

\[C(t)=\frac{k_BT}{\kappa}e^{-|t|/\tau},\]
\[f_c=\frac{\kappa}{2\pi\zeta},\]
\[S_x^{(1)}(f)=\frac{4k_BT\zeta}{\kappa^2+(2\pi f\zeta)^2},\quad f\ge0 .\]

The stated spectrum is one-sided in ordinary frequency. Its integral from zero to infinity equals \(C(0)\); doubling another time would double the variance. In an unbounded fluid with viscosity 1.00 mPa s, Stokes drag is \(6\pi\eta a=7.54\times10^{-10}\) N s/m. The default predicts \(f_c=60.5\) Hz and rms displacement 120 nm. The latter is roughly 0.18 waist, so the outer part of a long thermal trajectory may sample appreciable anharmonicity. A finite-depth well in an unbounded bath does not have a globally normalisable trapped equilibrium. These local equilibrium predictions apply to a sufficiently long-lived, locally sampled trapped state.

If a voltage detector gives \(V=gx\), the spectral corner and normalised autocorrelation time do not depend on the displacement gain \(g\). They determine only \(\kappa/\zeta\). Equipartition using a calibrated spatial variance gives \(\kappa\) independently of drag, provided mechanical fluctuations are in equilibrium and sufficiently resolved. If the viscosity or nearby-wall drag is uncertain, combining a known stage oscillation with the Brownian spectrum can determine further calibration quantities. Keep a stage-driven coherent peak out of the noise integral used for equipartition.

Account for exposure, sample size and unobserved motion

A camera typically reports the position averaged during an exposure of duration \(t_e\). Its variance is the double integral of the covariance over the exposure interval divided by \(t_e^2\). Evaluating the exponential local model and introducing \(u=t_e/\tau\) gives a multiplicative factor \(2(u-1+e^{-u})/u^2\). At \(u=1\) the reported physical-motion variance is only 0.736 of the instantaneous variance. Interpreting it as the full variance would overestimate stiffness by approximately 36 percent. Independent pixel-localisation noise adds additional variance with the opposite sign of bias.

Estimate that localisation noise with an effectively immobilised reference bead at a comparable brightness. Test the averaging correction by changing exposure duration at constant sampling frequency where possible. A camera that reports overlapping exposures has correlated measurement errors and needs a different covariance model. A long series of closely spaced frames also contains fewer independent thermal configurations than its frame count suggests. Compare estimates across several blocks substantially longer than the relaxation time; residual low-frequency drift can dominate even when high-frequency position precision looks excellent.

A separate check uses a known uniform liquid flow. At a steady small displacement \(x_s\), force balance gives

\[\kappa x_s=\zeta u_{\rm flow}.\]

For the default drag and stiffness a flow of 10 micrometres per second gives an offset of 26.3 nm. Repeat with reversed flow and use half the displacement difference to cancel an unknown zero-force position. Near a wall, the local flow at the bead and wall-corrected drag may differ from a stage velocity and the unbounded Stokes value. Those effects alter the mechanical calibration even if the optical field is unchanged.

A known stage oscillation calibrates position in a fluid

Brownian fluctuations alone determine the product of several unknown calibration parameters. A weak sinusoidal translation of the sample cell adds a known mechanical signal at a chosen frequency. Work in laboratory coordinates with a fixed optical trap and with fluid moving uniformly at velocity \(u(t)\) near the bead, well away from walls in the idealised model. The drag is now \(-\zeta(\dot x-u)\). Superposing a stage displacement \(A\cos\omega t\) on thermal fluctuations gives a deterministic response whose complex amplitude satisfies \((\kappa-i\omega\zeta)\widetilde x=-i\omega\zeta A\) in the negative-exponential convention. Therefore its amplitude ratio is

\[\frac{|\widetilde x|}{A}=\frac{\omega\tau}{\sqrt{1+\omega^2\tau^2}} .\]

At the trap corner the bead displacement amplitude is the known stage amplitude divided by the square root of two. A calibrated 80.0 nm stage oscillation there produces a predicted bead amplitude of 56.6 nm. If a position-sensitive receiver records a coherent peak amplitude of 0.170 V, its displacement gain is 3.01 mV/nm. At very low drive frequencies the fluid velocity is small for fixed stage amplitude and the trap holds the bead nearly stationary. At high enough frequency the bead follows the fluid displacement within the overdamped model, but finite hydrodynamic, optical and electronic bandwidths eventually matter.

After determining the gain, convert the surrounding voltage spectrum into spatial units and use the area of the stochastic fluctuation spectrum to find stiffness. Exclude the calibrated line and account for the finite spectral averaging window; a narrow drive that leaks into neighbouring bins can inflate the inferred thermal variance. Dividing measured stiffness by the independently determined corner rate then gives drag, which can be compared with the known viscosity and sphere radius. This avoids assuming free-space Stokes drag in the first step of a near-wall measurement.

There are two limiting systematic effects that a good fit to a single drive frequency cannot resolve. A frequency-dependent stage calibration changes the actual fluid displacement, and motion of the optics relative to the cell generates a direct detector background even with the bead immobilised. Repeat the drive above and below the corner, compare coherent phase as well as amplitude and acquire a trapped and immobilised reference. Correct conversion should explain the entire complex response with one spatial gain and one corner time. A single calibration number obtained at one frequency can hide a phase lag in the stage controller or in the optical position channel.

One last cross-check reverses the stage drive phase by half a period. The coherent bead response must reverse, while thermal background power remains unchanged. A component that fails this reversal could come from ambient vibration unrelated to the stage. Averaging the two traces with and without a sign change separates these contributions before either is used in an optical-force calibration.

Tutorial · Separate a unit-conversion error from excess noise

In a single-axis trap at 298 K a calibrated detector reports a variance of 2500 square nanometres after removal of localisation noise and exposure correction. An independent spectral fit gives a corner of 175 Hz. Infer stiffness, drag and the equivalent unbounded hydrodynamic radius for a liquid of viscosity 1.00 mPa s. Compare against a nominal 80 nm radius bead.

Infer the three quantities and test their consistency

A nanometre squared is \(10^{-18}\) square metres, so the variance is \(2.50\times10^{-15}\) square metres. Equipartition gives \(1.65\times10^{-6}\) N/m. Dividing by \(2\pi f_c\) gives drag \(1.50\times10^{-9}\) N s/m and radius \(\zeta/(6\pi\eta)=79.5\) nm. The agreement supports the local overdamped description under the assumed temperature and viscosity. Forgetting the factor of \(2\pi\) would predict a radius roughly six times larger and could be mistaken for a huge near-wall correction.

Now imagine that additional uncorrelated read noise had been left in the variance. This raises apparent variance, lowers the equipartition stiffness and would lower the inferred drag at the same fitted corner. Nearby-wall drag for a given radius instead normally raises hydrodynamic resistance for parallel translation. Their effects on the inferred radius thus differ in sign under the stated assumptions. Varying height and independently calibrating the measurement noise are appropriate diagnostics before attributing a discrepancy to bead dimensions.

Tutorial · Test a force-law inference with power and concentration controls

At fixed bead radius a trap is operated at two powers differing by a factor of two. Measured corner frequencies increase from 42 to 84 Hz, but rms displacement falls only from 110 to 87 nm. What does the harmonic model with fixed drag and temperature predict? Estimate a power-independent additive position-noise variance that could reconcile the two readings.

Use two variances to separate scale and offset

The doubled corner gives doubled stiffness if drag is unchanged. Physical thermal variance should halve. The predicted rms at higher power would be 77.8 nm if the lower-power value were entirely physical. Let the low-power measured variance be \(v+b\) and the high-power one \(v/2+b\), where \(b\) is a constant additive noise variance. Solving from 12,100 and 7,569 square nanometres gives \(v=9062\) and \(b=3038\) square nanometres, or approximately 55.1 nm rms noise.

This is a large inferred detector noise and demands an independent immobilised reference check. Power-dependent heating of the liquid changes temperature and viscosity, so an unchanged product inferred from only two quantities cannot prove that heating is absent. Test several powers, reverse their order to reveal slow drift and vary bead concentration to exclude overlapping particles in the tracking window. The inferred additive term should remain constant across those controls if it comes from the specified detector model.

Tolić-Nørrelykke and colleagues · position and force calibration using thermal and driven measurements