Where a thermoelectric cooler rejects its heat

A detector has to operate ten kelvin below the laboratory air temperature. A thermoelectric module beneath it can transport heat against that temperature difference when an electric current flows. Increasing the current also generates heat inside the module and warms its heat sink. Build an energy balance that predicts when additional current helps, when it costs more than it delivers and why a nominal cooling-capacity figure alone cannot specify an operating point.

Interactive calculation · Cooler capacity

Change the physical and numerical inputs of the notebook model. Compare the updated curves and balance checks with the recorded notebook figures. Each plot uses the same equations and assumptions documented in that notebook; its readout names any analytical or reduced-model reference.

Download this example’s editable notebook

Interactive calculation · Cooler cop

Change the physical and numerical inputs of the notebook model. Compare the updated curves and balance checks with the recorded notebook figures. Each plot uses the same equations and assumptions documented in that notebook; its readout names any analytical or reduced-model reference.

Download this example’s editable notebook

Save notebook (.ipynb)

Python runs on this device. The first run loads the scientific libraries.

The optional OneSimulation calculation uses the local OneSimulation installation. Save this notebook to run that comparison in Jupyter.

A temperature-controlled sensor is colder than its laboratory surroundings. Increase the current in a thermoelectric module and predict how the available cooling capacity changes. The model treats the effective properties of the entire module as constant. Edit the inputs and rerun the notebook.

Close the hot-side balance before computing cold-side capacity

Use the convective sink relation \(T_h=T_a+R_s Q_h\) to eliminate \(Q_h\). Heat transport at the hot face contains Peltier heat, half of the electrical dissipation and heat conducted back toward the cold plate. Current is positive in the cooling direction.

Compare the current sweep for two heat sinks

The coefficient of performance is plotted only where the module removes heat and consumes positive electrical power. A passive thermal leak at zero current can therefore appear below zero on the capacity panel.

Decide which operating point serves the measurement

Suppose the sensor dissipates 15 W. Use the arrays to find its lowest feasible drive current on the grid and compare the cooling margin with the point of maximum capacity. Repeat with an ambient temperature 5 K higher. Check whether the hot-face temperature remains within the range where your assumed constant properties are credible.

Resolve a cooling threshold between sweep points

A sampled current sweep can miss the exact point at which a heat load is first served. Use the rising flank as a bracket for a scalar solve and check the remaining heat-balance error. A local OneSimulation installation supplies its monotone inverse. The browser performs bisection with the same limits.